P: Polarizing filter.Used for polarizing testing of phoriz,Binocular balance of stereoscopic vision.
RMV: Red Maddox rod is set vertically.
RMH: Red Maddox rod is set horizontally.
WMV: White Maddox rod is set vertically.
WMH: White Maddox rod is set horizontally.
RL: Red lens.
GL: Green lens. +.12:+.12diopter spherical lens.Spherical power can be set by 0.12 diopter.
PH: A 1mm diameter pinhole is provided.This lens is used to determine if Poor vision is due to refractive abnormality or other caused.
6 U: 6 diopter base up prism.Used for the tests such as horizontal phoria test.
10 I 10 diopter base in prism.Used for the tests such as vertical phoria test.
+-.50: +-0.5 diopter cross cylinder with horizontal plus(+) axis.Used for presbyopia test.
OC: Occuluder
Item and range of measurement:
Myopia: 0 to -19.00 diopter in 0.25 diopter step (0.12 diopter step)
Hyperopia: 0 to +16.75 diopter in 0.25 diopter step (0.12 diopter step)
Astigmatism: 0 to -6.00 diopter(-8.00 diopter) in 0.25 diopter step (0.12 diopter step)
Phoria and vergency: 20 to 0 to 20 by 1 step;Inner 40;Outer 40(50);Upward and downward 40(46)
Near vision test: Main optical axes of both lenses converges by means of a tilt mechanism at 40cm in front of the eyes at 64mm P.D. If P.D. for far vision is less then 54mm,converging operation stops.
Binocular balance test: Rotary prism and polarizing lens methods are available.
P.D.: 48 to 75mm by 1mm step
Level adjustment: More than 4.5 for right and left banks